beautypg.com

Bio-Rad ChromLab™ Software User Manual

Page 297

background image

Using the C-96 Autosampler with NGC Systems

Instrument Guide

| 295

Setting Up the C-96 Autosampler for Use in ChromLab
Methods

This section briefly explains how to set up the autosampler for use in a ChromLab

method. For information about creating methods in ChromLab, see the NGC

Chromatography Systems and ChromLab Software User Guide.

To set up a method that uses the autosampler

1.

In ChromLab create a method that includes the autosampler in the fluidic
scheme. The Sample Application phase automatically displays Inject Using
Autosampler.

2.

In the Sample Application phase, increase the injection volume to include the
total volume of the installed autosampler loop and the tubing needed to
connect Port 6 on the autosampler to the Loop F port on the sample inject
valve.

Tip:

Typically, a sample loading volume of 2–3 ml should be sufficient to

completely load the entire sample in the autosampler loop onto the

column.

3.

Use the front panel on the C-96 autosampler to create a method.

See the C-96 Autosampler for BioLogic DuoFlow™ Chromatography System

Quick Start Guide or the C-96 Autosampler for BioLogic DuoFlow

Chromatography System User Manual for more information.

4.

Start the autosampler method and place the autosampler in REMOTE mode.

5.

In ChromLab, complete the method setup and start the run.

For multiple samples, perform a multi-run of the ChromLab method. The number of

runs (in the multi-run) should correspond to the total number of injections (from the

same vial or multiple vials) to be programmed on the autosampler.