beautypg.com

Bio-Rad ChromLab™ Software User Manual

Page 256

background image

B

| Troubleshooting the Instrument

254

| NGC Chromatography Systems and ChromLab Software

The UV baseline is
unstable or noisy
(continued).

The UV baseline shows
a reproducible zig-zag
or sawtooth trace
(continued).

UV or UV/Vis lamps are dirty or are
close to the end of their usable
life.

Check the lamp usage values in
the System Information dialog
box:

1. From the System Control tab
in ChromLab open File > System
Information > Detector tab.

2. Click Get Lamp Status.

Use the lamp usage values
(close to 2,000 hours for the
deuterium and tungsten
lamps; close to 5,000 hours or
100% PWM for LEDs), to
determine if the lamps are
nearing their end of life.

To change the UV flow cell see

Replacing the UV Flow Cell on
page

196

.

To change the lamps:

See

Replacing the Single-

Wavelength UV Detector LED
on page

201

See

Replacing the Multi-

Wavelength UV/Vis Detector
Lamps on page

209

Note:

For more information

about the Detector tab, see
the chapter System Control
in the NGC
Chromatography Systems
and ChromLab Software
User Guide.

Excessive pump
pulsations exhibit as
regular noise on the
baseline.

Air bubbles trapped in the pump
heads can produce exaggerated
pulsations, which appear as noisy
UV baselines.

Purge the pump heads to
remove bubbles.

Symptom

Possible Cause

Possible Solution