Inlet valve leds – Bio-Rad ChromLab™ Software User Manual
Page 51

Valves
Instrument Guide
| 49
Inlet Valve LEDs
The inlet valve modules are fitted with LEDs.
Solid green LEDs indicate the flow is from the system pump.
Solid blue LEDs indicate the flow is from the sample pump.
Blinking green LEDs indicate the line to plumb using the Point-to-Plumb
feature.
The Outlet LED is lit when the valve is in use.
When one sample inlet valve is in line, a single LED indicates the active
inlet port.
When two sample inlet valves are in line and sample flows from S2:
Both Outlet LEDs are lit.
The LED on the active port on S2 is lit.
The 8/BB on S1 is lit.
Two sample inlet valves (S1 and S2) in fluidic scheme
Valve
Port
Function
S1 and
S2
1–7
Sample inlets
S1
S2
8/BB
Inlet from sample inlet valve 2
Sample inlet
S1
S2
Out
Outlet to sample pump
Outlet to S1