Other Components
Instrument Guide
| 239
4.
Remove the side mount cover from the side of the instrument on which you
plan to attach the touch screen.
Note:
The top section in this image is a third tier. (The NGC Discover™
system is a 3-tier system.) The touch screen is attached at the top of the
second tier. If you have an NGC Quest™ or an NGC Scout™ system, you
will not have a third tier. Locate the side mount cover at the top of the
second tier.
5.
Remove the screws that secure the right and left side cable trough covers to
the back of the instrument.
Tip:
Set these components aside. You will use them in the steps that
follow.