beautypg.com

Set and activate acceptance limits – Bird Technologies SA-6000EX User Manual

Page 82

background image

Bird Technologies

66

Set and Activate Acceptance Limits

The limit line helps to set off those parts of a trace that are outside of
the acceptance (limit) values. It appears as a horizontal line at the
limit line value.

You can set an acceptance limit line to help visualize what part of the
measurement trace is not acceptable. The part of the trace that is not
acceptable will display in red. To view the limit line, you must have
the limit test ON. This feature is available only in the Measure Match
mode and the Fault Location mode.
1. If needed, turn the unit on and select the Measure Match mode

(refer to Figure 3 on page 18).

2. If needed, select the frequency (refer to Figure 4 on page 19 or

Figure 5 on page 20)

3. Make sure that the instrument is calibrated (refer to Calibration,

page 15).

4. Select the Fault Location mode (refer to Select Fault Location

Mode, page 48).

5. Follow the numbered steps in Figure 32, page 67 to turn ON the

limit test and set the limit line value.

Note:

You can set a limit line value with the limit test

turned OFF, but you will not see the line displayed.

This manual is related to the following products: