Setting the Maximum Distance
Note:
Changing the frequency span or the max distance
will automatically turn calibration off. Always set the
frequency span or max distance before calibrating the unit.
1. Select the Fault Location mode (Figure 19 on page 48).
2. Follow the numbered steps in Figure 21 on page 52 to enter the
maximum distance value.
3. At the warning message (Figure 22, page 53) press Enter to accept
your entry or Esc to cancel your entry. When you press Enter, the
Site Analyzer will display the Calibration screen so you can begin
the calibration process.
Figure 21 Setting the Maximum Distance
Step
Action
1
Press the Config key
2
Press the Distance softkey
3
Press the right or left arrow keys to scroll to Max
4
Numeric key pad - enter a maximum distance value
5
Press the Enter key after entering a value
6
Press the Esc key to exit the configuration screen