Recall a Saved Trace

When you recall a saved trace, it is displayed along with the trace cur-
rently being measured (two traces on the screen). To remove a recalled
trace from the display, press the Save/Recall softkey then press the
Esc key. With a current and recalled trace, the display behaves as fol-
lows:

The frequency range will change to the settings of the recalled
trace.

Recalled traces adopt the display type of the current trace
(envelope, maximum, minimum).

The resolution of the recalled trace does not affect the resolu-
tion of the current trace.

Markers always follow the current trace not the recalled trace.

When comparing the current trace to a recalled trace, each
data point on the current trace is compared to only the first
data point on the recalled trace (at the frequency of the
marker).

The status box at the top of the screen displays the resolution
and display type of the current trace on the top line and the
label text of the recalled trace on the bottom line.

Note:

Calibration is automatically turned off if the

recalled trace has a different frequency range from the trace
currently being measured.

The recalled a trace is drawn using the same display type (envelope,
maximum, minimum) as the trace currently being measured.

You must be in either Measure Match mode or Fault Location mode to
recall a saved trace.
1. Press the Save/Recall softkey
2. Follow the numbered steps in Figure 51, page 92 to select and

recall a saved trace.