beautypg.com

Select the measure match mode, Figure 3 selecting the measure match mode, Setting the measurement frequency – Bird Technologies SA-6000EX User Manual

Page 34

background image

Bird Technologies

18

Select the Measure Match Mode

Follow the numbered steps in Figure 3 to turn ON the Site Analyzer
and select the Measure Match mode.

Figure 3 Selecting the Measure Match Mode

Setting the Measurement Frequency

You must be in the Measure Match mode to set a frequency.

Frequencies can be set manually or chosen from a list of presets (band
list). If the start, stop, center, or span frequency is manually set, the
band list will indicate “Custom”. Be sure to set the frequency to a
value that is well beyond the normal range of the antenna. For exam-
ple, for a 450 MHz antenna, set the start frequency at 400 and the
stop frequency at 500 MHz, for an 800 MHz antenna, set the start fre-
quency at 700 and the stop frequency at 1,100 MHz.

Note:

Changing the frequency settings will automatically

turn calibration off. Always set the frequency before calibrating
the unit.

Step

Action

1

Turn On the Site Analyzer. Press ON/
OFF

key for ½ second.

2

Press the MODE key

3

Press the Measure Match key

This manual is related to the following products: