15

Chapter 2

Calibration

Calibration Accessories

Calibration Combination (Cal Combo) which contains:

One 50 ohm load

One Open standard

One Short standard

Test Cable (optional) – A phase-stable cable will provide con-
sistent, reliable results.

Note:

When using a test cable connected to the Site Ana-

lyzer, attach the Cal Combo to the end of the cable during cali-
bration.

Calibrating

For best results, set the frequency and calibrate the Bird Site Ana-
lyzer immediately before taking measurements.

Follow the numbered steps in Figure 2, page 16 to calibrate the Site
Analyzer.

Note:

When calibrated, the Site Analyzer’s main screen

will display “Calibration: FULL”. When uncalibrated, the Site
Analyzer’s main screen will display “Calibration: OFF”.

Note:

Calibration is performed at the highest number of

data points (949 points) and the calibration graphs are dis-
played using the currently selected display method (envelope,
maximum, or minimum).