Calibrate the site analyzer, Select the measurement units – Bird Technologies SA-6000EX User Manual
Page 37

Measure Match Mode
21
Calibrate the Site Analyzer
If you need to use an extension cable to connect the Site Analyzer to
the system, connect the extension cable to the antenna test port on the
Site Analyzer and then connect the Bird Cal Combo to the end of the
extension cable. Follow the calibration procedure in
Chapter 2, Calibration, page 15.
Note:
When using an extension cable, you should use a
phase stable cable to ensure performance and accuracy. Bird
offers optional phase stable cables that will not degrade the per-
formance of your Site Analyzer. Using a non-Bird phase stable
cable may decrease and degrade the accuracy of the system.
Select the Measurement Units
You can select units before or after making a measurement. When
making a match measurement, you can select one of three units for
the Y (vertical) scale,
•
VSWR (ratio)
•
Rtn Loss (dB)
•
Cbl Loss (dB)
1. Make sure that the instrument is turned on and in the Measure
Match mode (refer to Select the Measure Match Mode, page 18)
2. Select Frequency (refer to Setting the Measurement Frequency,
3. Make sure that the instrument is calibrated (refer to Calibration,
4. Follow the numbered steps in Figure 6, page 22 to set
measurement units.