beautypg.com

Calculate:ndb:bandwidth|bwidth, Calculate:normalize:state, Calculate:ocbw:state – GW Instek GSP-930 Programming Manual User Manual

Page 92

background image

GSP-930 Programming Manual

90

Example

:CALC:NDB:STAT ON

:CALCulate:NDB:BANDwidth|BWIDth?

Query

Description

Returns the NdB bandwidth measurement.

Query Syntax

:CALCulate:NDB:BANDwidth|BWIDth?

Return parameter

NdB bandwidth in Hz.

Example

:CALC:NDB:BAND?
>5.5e+04

:CALCulate:NORMalize:STATe

Set

Query

Description

Turns the tracking generator normalization on/off
or queries its state.

Syntax

:CALCulate:NORMalize:STATe{ON|OFF|1|0}

Query Syntax

:CALCulate:NORMalize:STATe?

Parameter

0

Turn normalization off.

1

Turn normalization on.

OFF

Turn normalization off.

ON

Turn normalization on.

Return parameter 0

normalization is off.

1

normalization is on.

Example

:CALC:NORM:STAT ON

:CALCulate:OCBW:STATe

Set

Query

Description

Turns the OCBW measurement on/off or queries
its state.

Syntax

:CALCulate:OCBW:STATe {ON|OFF|1|0}

Query Syntax

:CALCulate:OCBW:STATe?

Parameter

0

Turn OCBW off.

1

Turn OCBW on.

OFF

Turn OCBW off.