GSP-930 Programming Manual

152

Query Syntax

[:SENSe]:SEMask:OFFSet:STARt:RELative?

Parameter/
Return parameter

dBc

offset 1~5

Example

SEM:OFFS1:STAR:REL 2.5e+1

[:SENSe]:SEMask:OFFSet:STATe

Set

Query

Description

Turns the selected offset on/off or queries its state.

Syntax

[:SENSe]:SEMask:OFFSet:STATe {OFF|ON|0|1}

Query Syntax

[:SENSe]:SEMask:OFFSet:STATe?

Parameter

0

Turns the selected offset off.

1

Turns the selected offset on.

OFF

Turns the selected offset off.

ON

Turns the selected offset on.

Return parameter 0

The selected offset is off.

1

The selected offset is on.

Example

SEM:OFFS1:STAT 1

[:SENSe]:SEMask:OFFSet:STOP
:ABSolute

Set

Query

Description

Sets or queries the amplitude of the stop frequency
of the Absolute Mask for the selected offset.

Syntax

[:SENSe]:SEMask:OFFSet:STOP:ABSolute

Query Syntax

[:SENSe]:SEMask:OFFSet:STOP:ABSolute?

Parameter/
Return parameter

dBm

offset 1~5

Example

SEM:OFFS1:STOP:ABS 1.5e+1

[:SENSe]:SEMask:OFFSet:STOP:
ABSolute:COUPle

Set

Query