Operation status register, Questionable status register – GW Instek GSP-930 Programming Manual User Manual
Page 47

REMOTE CONTROL
45
Operation Status Register
Overview
The Operation Status Register Group indicates
the operating status of the GSP-930.
Bit Summary
Bit Weight Description
3
8
Sweeping: Indicates that a sweep is in
progress.
4
16
Measuring: The instrument is currently
performing a measurement.
5
32
Waiting for Trigger: The instrument is in
a “wait for trigger” state.
Questionable Status Register
Overview
The Questionable Status Register Group
indicates if any limits have been tripped.
Bit Summary
Bit Weight Description
5
32
Frequency Status Summary Bit: This is
the summary bit of the Frequency Status
Register.
8
256
Uncal: This bit is set when a signal level
occurs because the sweep is too fast.
9
512
Limit fail: This bit is set to 1 when the
limit line has been violated.
10
1024
ACP Limit Status Summary Bit: This is
the summary bit for the ACP Limit
Status Register.
11
2048
SEM Limit Status Summary Bit: This is
the summary bit for the SEM Limit
Status Register.
12
4096
TOI Limit Status Summary Bit: This is
the summary bit for the TOI Limit
Status Register.
13
8192
Pmet Limit Fail: This bit is set to 1 when
the power meter limit has been violated.