beautypg.com

Calculate:limit:marker, Calculate:limit:pass:mode, Calculate:limit:state – GW Instek GSP-930 Programming Manual User Manual

Page 80

background image

GSP-930 Programming Manual

78

:CALCulate:LIMit:MARKer

Set

Description

Sets the current marker position to a point on a
limit line. The vertical position of the point is the
marker’s vertical position + a user-defined offset.

Syntax

:CALCulate:LIMit:MARKer ,

Parameter

point 1~10

dB

Seleted limit line

Example

:CALC:LIM1:MARK 5, 20

:CALCulate:LIMit:PASS:MODE

Set

Query

Description

Sets or queries the Pass/Fail conditions for the
limit line testing.

Syntax

:CALCulate:LIMit:PASS:MODE {ALL|MAX|MIN}

Query Syntax

:CALCulate:LIMit:PASS:MODE?

Parameter/
Return Parameter

ALL

All-in.

MAX

Max-In

MIN

Min-In

Example

:CALC:LIM:PASS:MODE ALL

:CALCulate:LIMit:STATe

Set

Query

Description

Turns the limit line Pass/Fail test on/off.

Syntax

:CALCulate:LIMit:STATe {ON|OFF|1|0}

Query Syntax

:CALCulate:LIMit:STATe?

Parameter

0

Pass/Fail test is off.

1

Pass/Fail test is on.

OFF

Pass/Fail test is off.

ON

Pass/Fail test is on.

Return parameter 0

Pass/Fail test is off.

1

Pass/Fail test is on.