Calculate:demod:reset, Calculate:jitter:state, Calculate:jitter:carrier:power – GW Instek GSP-930 Programming Manual User Manual

Page 77

background image

REMOTE CONTROL

75

Return parameter 0

FM Analysis is off.

1

FM Analysis is on.

Example

:CALC:DEM:FM:STAT 1

:CALCulate:DEMod:RESet

Set

Description

This command will reset the max and min records
for the current demodulation analysis.

Syntax

:CALCulate:DEMod:RESet

:CALCulate:JITTer:STATe

Set

Query

Description

Sets or queries the state of the Jitter Analysis
function.

Syntax

:CALCulate:JITTer:STATe {ON|OFF|1|0}

Query Syntax

:CALCulate:JITTer:STATe?

Parameter

0

Turn Jitter Analysis off.

1

Turn Jitter Analysis on.

OFF

Turn Jitter Analysis off.

ON

Turn Jitter Analysis on.

Return parameter 0

Jitter Analysis is off.

1

Jitter Analysis is on.

Example

:CALCulate:JITTer:STATe 1

:CALCulate:JITTer:CARRier:POWer?

Query

Description

Returns the carrier power in the current Y-axis
units.

Query Syntax

:CALCulate:JITTer:CARRier:POWer?

Return parameter

In the current Y-axis units.

Query Example

:CALC:JITT:CARR:POW?
>-5.237e+01