REMOTE CONTROL

71

Parameter

Marker number.

Return parameter in dB.

Example

:CALC:DELT1:Y?
>-1.032e+1

:CALCulate:DEMod:AM:RESult:CURRent?

Query

Description

Returns the current measurement results for AM
demodulation as a comma separated string.

Query Syntax

:CALCulate:DEMod:AM:RESult:CURRent?

Return parameter

depth

Modulation depth in %.

rate

Modulation rate in Hz.

power

Carrier power in the current Y-axis units.

offset

Carrier frequency offset in Hz.

sinad

Signal to noise and distortion ratio in dB

Query Example

:CALC:DEM:AM:RES:CURR?
>9.840e+1,1.02e+2,-1.12e+1,3.21e+1,1.61e+1

:CALCulate:DEMod:AM:RESult:MINimum?

Query

Description

Returns the minimum recorded measurement
results for AM demodulation as a comma
separated string.

Query Syntax

:CALCulate:DEMod:AM:RESult:MINimum?

Return parameter

depth

Modulation depth in %.

rate

Modulation rate in Hz.

power

Carrier power in the current Y-axis units.

offset

Carrier frequency offset in Hz.

sinad

Signal to noise and distortion ratio in dB

Query Example

:CALC:DEM:AM:RES:MIN?
>9.840e+1,1.02e+2,-1.12e+1,3.21e+1,1.61e+1