beautypg.com

Sense]:acpr:llimit, Sense]:acpr:help:state, Sense]:acpr:space – GW Instek GSP-930 Programming Manual User Manual

Page 125

background image

REMOTE CONTROL

123

Parameter

power or voltage

Return parameter

Example

:ACPR: HLIM -3.0e+1

[:SENSe]:ACPR:LLIMit

Set

Query

Description

Sets or queries the low limit for the main channel.
Used with ACPR measurement.

Syntax

[:SENSe]:ACPR:LLIMit

Query Syntax

[:SENSe]:ACPR:LLIMit?

Parameter

power or voltage

Return parameter

Example

:ACPR:ACH1:LLIM -5.0e+1

[:SENSe]:ACPR:HELP:STATe

Set

Description

Turns the on-screen help on/off.

Syntax

[:SENSe]:ACPR:HELP:STATe {OFF|ON|0|1}

Parameter

0

Turn help off.

1

Turn help on.

OFF

Turn help off.

ON

Turn help on.

Example

:ACPR:HELP:STAT ON

[:SENSe]:ACPR:SPACe

Set

Query

Description

Sets or queries the channel spacing between the
main channels.

Syntax

[:SENSe]:ACPR:SPACe

Query Syntax

[:SENSe]:ACPR:SPACe?