beautypg.com

Sense]:average:state, Sense]:average:type, Sense]:bandwidth|bwidth[:resolution – GW Instek GSP-930 Programming Manual User Manual

Page 128

background image

GSP-930 Programming Manual

126

Parameter/
Return parameter

Example

:AVER:COUN 20

[:SENSe]:AVERage:STATe

Set

Query

Description

Turns the Average function on/off or queries its
state.

Syntax

[:SENSe]:AVERage:STATe {OFF|ON|0|1}

Query Syntax

[:SENSe]:AVERage:STATe?

Parameter

0

Turn the Average function off.

1

Turn the Average function on.

OFF

Turn the Average function off.

ON

Turn the Average function on.

Return parameter 0

The Average function is off.

1

The Average function is on.

Example

:AVER:STAT ON

[:SENSe]:AVERage:TYPE

Set

Description

Sets the method that the Average function uses to
calculate the average.

Syntax

[:SENSe]:AVERage:TYPE
{VOLTage|LOGarithmic|POWer}

Query syntax

[:SENSe]:AVERage:TYPE?

Parameter/Return
parameter

VOLTage

Sets Average to voltage

LOGarithmic Sets Average to logarithmic

POWer

Sets Average to power

Example

:AVER:TYPE VOLT

[:SENSe]:BANDwidth|BWIDth[:RESolution]

Set

Query

Description

Sets or queries the resolution bandwidth (RBW).