beautypg.com

Sense]:sweep:egate:state, Sense]:sweep:time, Sense]:sweep:time:auto – GW Instek GSP-930 Programming Manual User Manual

Page 162

background image

GSP-930 Programming Manual

160

Syntax

[:SENSe]:SWEep:EGATe:LENGth

Query Syntax

[:SENSe]:SWEep:EGATe:LENGth?

Parameter/
Return parameter

Gate length time in seconds

Example

SWE:EGAT:LENG 10 ms

[:SENSe]:SWEep:EGATe:STATe

Set

Query

Description

Turns the gated sweep mode on/off or queries its
state.

Syntax

[:SENSe]:SWEep:EGATe:STATe {OFF|ON|0|1}

Query Syntax

[:SENSe]:SWEep:EGATe:STATe?

Parameter

0

Turns gated sweep mode off.

1

Turns gated sweep mode on.

OFF

Turns gated sweep mode off.

ON

Turns gated sweep mode on.

Return parameter 0

Gated sweep mode is off.

1

Gated sweep mode is on.

Example

:SWE:EGAT:STAT 1

[:SENSe]:SWEep:TIME

Set

Query

Description

Sets the sweep time.

Syntax

[:SENSe]:SWEep:TIME

Query Syntax

[:SENSe]:SWEep:TIME?

Parameter/
Return parameter

Sweep time in seconds

Example

SWE:TIME 60 ms

[:SENSe]:SWEep:TIME:AUTO

Set

Query

Description

Turns the Sweep time setting to auto (on) or
manual (off).