REMOTE CONTROL

81

:CALCulate:MARKer:FCOunt:X?

Query

Description

Returns the counter frequency of the selected
marker in Hz.

Query Syntax

:CALCulate:MARKer:FCOunt:X?

Parameter

Marker number 1~6.

Return parameter

Frequency in Hz.

Example

:CALC:MARK1:FCO:X?
>2.0083e+8

:CALCulate:MARKer:NOISe:STATe

Set

Query

Description

Sets or queries the state of the Marker Noise
function.

Syntax

:CALCulate:MARKer:NOISe:STATe {ON|OFF|1|0}

Query Syntax

:CALCulate:MARKer:NOISe:STATe?

Parameter

Marker number 1~6.

0

Turn marker noise off.

1

Turn marker noise on.

OFF

Turn marker noise off.

ON

Turn marker noise on.

Return parameter 0

Marker noise is off.

1

Marker noise is on.

Example

:CALC:MARK2:NOIS:STAT ON

:CALCulate:MARKer:NOISe:Y?

Set

Query

Description

Returns the normalized noise level over a BW of
1Hz from the marker position.

Query Syntax

:CALCulate:MARKer:NOISe:Y?