5.3

Wiring and Piping Examples

5.3.1

Wiring and Piping for Automatic Calibration

Averaging Converter (AV550G)

Detector

Solenoid

Valve

Reference

Gas Line

Flowmeter

Needle Valve

Air Set

(Instrument Air)

Span Gas Cylinder

Power Supply

Zero Gas Cylinder

Stop Valve

Pressure Regulator

Solenoid

Valve

Flow

meter

Instrument Air

Calibration Gas Line

Needle Valve

: If optional 24 V outputs are specified for solenoid valves (Option Code 0/240), no external
power supply for solenoid valves is required. The solenoid valves are powered from the
AV550G Averaging Converter. Never connect external power sources in the wiring for
solenoid valves.

ء

ء

ء

ء

Figure 5.15 Typical Diagram for Automatic Calibration

5.3.2

Wiring and Piping for Automatic Calibration and 3rd Gas Indication Check

Averaging Converter (AV550G)

Analog Output

(Averaged and Individual Output)

Contact Output

Contact Input

Power Supply

Span Gas Cylinder

(Instrument Air)

Zero Gas Cylinder

Third Check Gas Cylinder

Pressure Regulator

Stop Valve

Solenoid
Valve

Flow
meter

Needle Valve

Calibration Gas Line

Air Set

Needle Valve

Flowmeter

Instrument
Air

Referece
Gas Line

Solenoid
Valve

Detector

: If optional 24 V outputs are specified for solenoid valves (Option Code 0/240), no external
power supply for solenoid valves is required. The solenoid valves are powered from the
AV550G Averaging Converter. Never connect external power sources in the wiring for
solenoid valves.

ء

ء

ء

ء

Figure 5.16 Typical Diagram for Automatic Calibration and 3rd Gas Indication

Check