beautypg.com

3 display configuration, 6 selection of gas to be measured, Measurement gas – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 116

background image

IM 11M12D01-01E

7-6

7.6 Selection of Gas to be Measured

Exhaust gas includes steam resulting from combustion of hydrogen and oxygen. You
may choose whether to display the oxygen concentration including the oxygen contained
in the steam, or display the computed "dry" oxygen concentration (a value compensated
for the oxygen in steam) which results in a higher concentration value. For details, refer
to " Fuel Settup" in Sec. 8.5.3

1) Touch the Setup key to switch to the [Execution/Setup] display.
2) Select [Commissioning] --> [Basic Setup] --> [Select Measurement gas] and the

[Measurement gas] display appears.

3) Touch the Enter key and the Measurement gas selection window opens for you to

select the measurement gas.

The Measurement gas selection applies to all channels, you cannot specify it on a
channel-by-channel basis.

Enter

Measurement gas

F7.5E.EPS

Select measurement gas:

Wet

Dry

See Manual for details

!

Fig. 7.5 Measurement gas selection