2-4
Terminal Box Case: Material; Aluminium alloy
Terminal Box Paint Color:
Case; Off-white (Munsell 0.6GY3.1/2.0)
Cover; Moss green (Munsell 2.5Y8.4/1.2)
Finish: Polyurethane corrosion-resistant coating
Gas Connection:
Rc1/4 or 1/4FNPT
Wiring Connection: G1/2, Pg13.5, M20 by 1.5 mm, 1/2 NPT
Installation:
Flange mounting
Probe Mounting Angle:
Horizontal to vertically downward.
When the probe insertion length is 2 m or less, can install at angles from
horizontal to vertically downward.
When the probe insertion length is 2.5 or more, mount vertically downward
(within
Ϯ5Њ), and if using a probe protector install at angles between horizontal
and vertically downward (within
Ϯ5Њ).
Weight:
Insertion length of 0.4 m:
approx. 6 kg (JIS 5K 65) / approx. 11 kg (ANSI 150 4)
Insertion length of 1.0 m:
approx. 8 kg (JIS 5K 65)) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.5 m:
approx. 10 kg (JIS 5K 65)) / approx. 15 kg (ANSI 150 4)
Insertion length of 2.0 m:
approx. 12 kg (JIS 5K 65)) / approx. 17 kg (ANSI 150 4)
Insertion length of 3.0 m:
approx. 15 kg (JIS 5K 65)) / approx. 20 kg (ANSI 150 4)
Insertion length of 3.6 m:
approx. 17 kg (JIS 5K 65)) / approx. 22 kg (ANSI 150 4)
Insertion length of 4.2 m:
approx. 19 kg (JIS 5K 65)) / approx. 24 kg (ANSI 150 4)
Insertion length of 4.8 m:
approx. 21 kg (JIS 5K 65)) / approx. 26 kg (ANSI 150 4)
Insertion length of 5.4 m:
approx. 23 kg (JIS 5K 65)) / approx. 28 kg (ANSI 150 4)