2. Specifications

2.3

High-Temperature Separate-type Detector and Related
Equipment

2.3.1

ZR22G (0.15m) High-Temperature Separate-type Detector

Standard Specifications

Construction: Water-resistant, non-explosionproof

Probe length: 0.15 m

Terminal box: Aluminium alloy

Probe material: Probe material in contact with gas: SUS 316 (JIS) (Probe), SUS 304

(JIS) (Flange), Zirconia (Sensor), Hastelloy B, (Inconel 600, 601)

Weight:

Approx. 3 kg

Installation:

Flange mounting (The use of high-temperature detector probe adapter

ZO21P-H is necessary.)

Flange standard: JIS 5 K 32 FF equivalent (thickness varies)

Mounting angle: Any angle between horizontal and vertical (high-temperature probe is

fitted with an adapter) Reference gas and calibration gas piping
connection:Rc 1/4 or 1/4 NPT female

Cable inlet:

G 1/2, Pg 13.5, M20

ϫ 15, 1/2 NPT

Ambient temperature: -20 to 150

ЊC

Sample gas temperature: 0 to 700

ЊC (temperature at the measuring point of the sam-

pling gas. 0 to 750

ЊC or 0 to 1400ЊC when the probe adapter for high

temperature is used.

Temperature of the probe adapter shall not exceed 300

ЊC to protect the

gasket and avoid the bolts seizing together.

Sample gas pressure: -0.5 to +5 kPa: when used at the range of more than 0 to

25 vol% O

2

, -0.5 to +0.5 kPa. (An auxiliary ejector is required for

negative pressure application.)

Model and Code

Refer to “Model and Code” in page 2-5.

External Dimensions

Refer to the Figure in page 2-6.