2 oxygen concentration alarm setting, 1 setting the alarm values (individual settings) – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 159

IM 11M12D01-01E
9-9
9. Calibration
Table 9.1 When Auto Calibration conflicts with Other Commands
T9.1E.eps
X : Cannot perform calibration
Wait : Wait until other command finishes before starting calibration
Purging
the
channel
Other
channel
CAL
Indication
Check in
other channel
Blowback in
other
channel
CAL chan.
in warmup
status
CAL chan.
in ERR
status
CAL chan.
now
disabled
Maintenance
Man. Cal.
SemiAuto
AutoCal
Wait
Wait
Wait
Wait
Wait
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
Other
commands
below.
Table 9.2
T9.2E.eps
Item
Initial value
Mode Manual
Points Span-Zero
Zero gas conc.
1.00%
Span gas conc.
21.00%
Hold time
3 min. 00 sec.
Cal. time
3 min. 00 sec.
Interval
1 days 00 hrs
Starting date
yy01mm01dd01
Starting time
hh00mm00