IM 11M12D01-01E

8-16


This setting example shows how to set "When contact input 1 Opens, Start Semi-Auto
Calibration".

1) From the panel basic display touch [Setup] key, and the [Execution/Setup] display

appears.

2) Select [Commissioning] and the Commissioning display appears.
3) Select [Contact Setup] then [Input Contacts].
4) Select [Input 1] . Function Selection window (Fig 8.16) appears.
5) Select [Calibration Start].
6) For [Input contact 1] select (operate if) [Closed]. (You can select Open or Closed).
7) Change [Closed] to [Open].


All contact inputs are set to [No function](disabled) prior factory shipment or after data
initialization..

Enter

F8.16E.EPS

Input 1 : Open

Input 1

Input 2

Input 2

Close

Input contacts

No function
Range change
Calibration start
lnd. check start
Blow back start
Cal. gas press. low
Process gas alarm

No function

Fig. 8.16 Contact Input Setting display