10-9

10. Other Functions

10.3 Other Display-related Functions

10.3.1

Auto-return Time

On the Execution/Setup display or lower level menu displays, if no keys are touched for
a preset time, the "auto-return" time, then the display will automatically revert to the
basic panel display. The "auto-return" time can be set in the range 0 to 255 minutes. If it
is set to 0, then the display does not automatically revert.


1) On the basic panel display, touch the Setup key to display the Execution/Setup

display, then select [Maintenance]. Select [Display setup] then [Auto return time].

2) The numerical entry display appears for you to enter the desired "auto-return" time.

To set an "auto-return" time of one hour, enter 060.

3) If you set 0, then the "auto-return" function does not operate.


The Auto return time is set to 0 prior factory shipment or after data initialization.

10.3.2

Selecting Language

You can select a display language from among English, Japanese, German and French.
The display language is set to the one specified in the purchase order when the analyzer
is shipped from the factory.


1) On the basic panel display, touch the Setup key to display the Execution/Setup

display.

2) Select [Maintenance], then [Display setup].
3) Select [Language] and the dropdown selection (Figure 10.7) allows you to select the

desired language.

Enter

F60E.EPS

Display setup

Trend graph

Auto return time: 0 min

Language:

Lcd auto 0 min

Display contrast

Tag name:

English

Japanese
Deutsch
Français

Figure 10.7 Selecting the Display Language