4 av550g averaging converter – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 31

IM 11M12D01-01E
2-12
2.4
AV550G Averaging Converter
2.4.1
Standard Specification
Compatibility of Detectors : ZR22G, ZO21D, ZO21DW
Number of Detectors : 1 to 8
Expandable up to 8
Bases, these for 4CH and 8CH are available
Averaging interval : 0.2 seconds
Display: 5.7 inches color LCD display of size 320 by 240 dot with touch screen
Output Signal: 4 to 20 mA DC (maximum load resistance 550
⍀)
Average-value Output; 3 points
Independent Output; Output to each channel
Common isolation / Individual isolation selectable
(Note) Number of averaging output is 2 when suffix code “ -F”
(FOUNDATION Fieldbus communication) is selected.
Independent Output; Output to each channel
Common isolation / Individual isolation selectable.
Used exclusively for communication when suffix code “ -F”
(FOUNDATION Fieldbus communication) is selected.
Contact Output: Contact capacity 30V DC 3A, 250V AC 3A (resistive load)
Normally open / normally closed selectable
Common Contact Output; 5 points, Four of the output points can be selected to either
normally energized or normally deenergized status.
Contact output 5 is normally energized.
Contact Output for Individual Channel Fail; Output to each channel
Normally energized.
Solenoid Valve Contact Output: Contact capacity 30V DC 1A, 250V AC 1A,
24V DC (option code "/24" ) maximum output current 50 mA
Contact Input: 2 points, voltage free contacts
Ambient Temperature: -5 to +50
ЊC
Storage Temperature: -20 to +70
ЊC
Humidity Range: 10 to 85%RH (non-condensing)
Installation Altitude: 2000 m or less
Category based on IEC 1010: II (Note)
Pollution degree based on IEC 1010:2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand
voltage. Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other
inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.
Power Supply Voltage: Ratings; 100 / 115 V AC, 230 V AC
Acceptable range; 85 to 126.5 V AC, 195.5 to 253 V AC
Power Supply Frequency: Rating; 50/60 Hz
Acceptable range; 50Hz
±
5%, 60Hz
±
5%
Power Consumption:
Max 1 kw for steady operation
Max 1.8 kw for warm-up