beautypg.com

6 other equipment – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 40

background image

IM 11M12D01-01E

2-21

2. Specifications

2.6

Other Equipment

2.6.1

Dust Filter for the Detector (Part No.: K9471UA)

This filter is used to protect the detector sensor from a corrosive dust component or from
a high concentration of dust when the oxygen concentration in utility boilers or concrete
kilns are to be measured.
This filter requires a measurement gas flow of 1m/sec or faster, for circulation of the gas
in the sensing part.

• Standard specification

Applicable detector: Standard-type detector for general-use (the measurement gas

flow should be approximately perpendicular to the probe.)

Mesh: 30 microns
Material: Sic (Filter), SUS316 (JIS)
Weight: Approx. 0.2 kg

T2.7E.EPS

Part No.

K9471UA

K9471UX

Description

Filter

Tool

51

32

10

SiC Caborundum filter

Increasing of insertion length

F2.8E.EPS

Unit : mm

2.6.2

Dust Guard Protector (Part No.: K9471UC)

Recommended to be used when sample gas is likely to flow directly into the cell due to
its flow direction in the stack or the like, flammable dust may go into the cell, or water
drops are likely to fall and remain in the cell during downtime or the like due to the
installation position.
Material: SUS316
Weight: Approx. 0.3 kg

4-

л

6

122

л

50.8

100

Increasing of insertion length

F11-1.eps