4 display functions, 7 current output range setting, Ma-output range – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 117

IM 11M12D01-01E
7-7
7. Startup
7.7 Current Output Range Setting
The output ranges (two ranges) of each channel are independently settable. Ranges are
switched by contact input. Contact OFF results in Range 1, and contact ON results in
range 2. For details, refer to Sec. 8.5 Setting Contact Inputs. Shipping time default is
for each range to be set to 0 to 25% O
2
. The ratio of the maximum range setting to the
minimum is restricted as described in Sec. 8.1 Setting Analog Outputs.
1) Touch the Setup key to switch to the [Execution/Setup] display.
2) From the [Commissioning] display select [mA-output setup], then [Set Range]. (See
Fig. 7.6).
3) Use [
᭣ ] and [᭤] keys to select the channel.
4) Set [Min.] O
2
concentration corresponding to 4 mA output, then set [Max.] O
2
concentration corresponding to 20 mA output (select, then touch Enter key).
5) The numeric data entry display shown in Fig. 7.7 appears; enter the desired oxygen
concentration setting. To set 10% O
2
, input [010] and touch Enter.
6) As necessary, repeat for each averaging output a, b and c of other channels.
Enter
mA-output range
max: 25 %O2
Range1
Range2
F7.6E.EPS
᭜
min: 0 %O2
᭜
max: 25 %O2
᭜
min: 0 %O2
Ch1
Enter
New value:
0
1
4
7
2
3
5
6
8
9
025 %O2
F17E.EPS
Fig. 7.6 Current-output range Fig. 7.7 Numerical entry