8. Detailed Data Setting


1) From "mA-output setup" on the Commissioning display, select "mA-outputs hold"

and touch the Enter key. A display like Fig. 8.9 is displayed.

2) Select an item you want to change, and touch the Enter key.
3) In the selection window that appears, select the desired Hold method and touch the

Enter key.

4) Select other items as desired and set as desired, as in the previous step. If you select

Preset Value then you can set a current output value.

5) Touch the Reject key to return to the "mA-outputs hold" display.
6) If you select Preset Value and touch the Enter key then the display shown in Fig. 8.10

appears.

7) Selecting a preset setting and touching the Enter key displays the numeric entry

display, allowing you to enter a preset current in the range 2.4 mA to 21.6 mA.

Enter

F8.9E.EPS

Maintenance: Non-Hold

Cal/Blow/Ind: Hold

Proc. gas alm: Hold

Error: Hold

Ave group invalid:

4mA

mA-outputs hold

Warm up: 4mA

Enter

F40E.EPS

Warm up: 4. 0 mA

Maintenance: 4. 0 mA

Cal/Blow/Ind: 4. 0 mA

Proc. gas alm: 4. 0 mA

Error: 4. 0 mA

Ave group invalid:

4. 0 mA

Preset vaiue

Fig. 8.9 Current output hold Fig. 8.10 Preset value setting


Output hold settings are set to defaults in Table 8.2 prior factory shipment or after data
initialization..

Tbale 8.2 Defaults for output hold and preset values

Status

Output hold value

Preset value

Warmup, Purging

4mA

4mA

(Under) Maintenance

Hold last value

4mA

(Under) Calibration
Indication Check
Blowback
Process Gas Alarm

Hold last value

4mA

Error

Hold last value

4mA

No Valid Channels in Averaging Group

4mA

4mA

Hold last value

4mA

T8.2Eeps