Manual ind. check – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 178

background image

IM 11M12D01-01E

10-16

F10.16E.EPS

Enter

21.00%

Ch2

1min./div.

0.98%

20.9 %O2

CHK. TIME

Manual ind. check

Channel No.

under calibration

Setting value of

span gas concentration

Setting value of

zero gas concentration

Actual measurement value

Fig. 10.16 Span Indication Check

9) If you touch the Enter key then span gas check ends and you are returned to the

Check gas selection display. Close the span gas flow control valve and tighten the
lock nut used to prevent it from loosening during measurement.

10) Select [Zero gas ind. check] and touch the Enter key.
11) Open the zero gas flow control valve, and adjust flow rate to 600

Ϯ60 ml/min.

12) Select [Start ind. check] and touch the Enter key. An oxygen concentration graph

like Figure 10.17 is displayed, and [CHK.TIME] status mark flashes. Wait for the
graph measured value trace to stabilize. Compare the measured zero gas concentra-
tion displayed at the bottom of the display against the certified concentration of zero
gas to check for any discrepancy or error.

F10.17E.EPS

Enter

21.00%

Ch2

1min./div.

0.98%

0.98 %O2

CHK. TIME

Manual ind. check

Fig. 10.17 Zero Indication Check

13) If you touch the Enter key then zero gas check ends and you are returned to the

Check gas selection display. Close the zero gas flow control valve and tighten the
lock nut used to prevent it from loosening during measurement.

14) Select [Third gas ind. check] and touch the Enter key.
15) Open the third gas flow control valve, and adjust flow rate to 600

Ϯ60 ml/min.

16) Select [Start ind. check] and touch the Enter key. An oxygen concentration graph

like Figure 10.18 is displayed, and [CHK.TIME] status mark flashes. Wait for the
graph measured value trace to stabilize. Compare the measured third gas concentra-
tion displayed at the lower left of the display against the certified concentration of
third gas to check for any discrepancy or error.

17) If you touch the Enter key then third gas check ends and you are returned to the

Check gas selection display. Close the third gas flow control valve and tighten the
lock nut used to prevent it from loosening during measurement.

18) Select [End]. An oxygen concentration graph (Figure 10.19) is displayed and hold

time begins. After the preset time has elapsed then the display reverts to channel
select. If you touch the Enter key during hold time then hold time is skipped.