6 starting auto indication check, 1 removing and attaching the front cover, 1 removing the front cover – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 212: 2 attaching the front cover, 2 hot swap function

IM 11M12D01-01E
12-2
12.1.2
Operations When an Error Occurs
When an error occurs, the averaging converter operates as follows.
When a channel-specific error occurs:
1) The power to the heater of the corresponding detector is removed.
2) On the multi-channel display, a red error code is indicated under the relevant channel
and an error icon flashes in the switch indication area (Figure 12.1).
3) On the single-channel display, a large error code is indicated in the oxygen concentra-
tion data area and an error icon flashes in the switch indication area if the display
shows the relevant channel (Figure 12.2). If not, only a flashing error icon is indi-
cated.
4) The relevant channel is dropped out of the averaging group. The averaging outputs
continue with exception of the channel.
5) The corresponding individual analog output is put into the status specified in the
Output Hold Setup display. No other individual analog outputs are affected.
6) The red lamp turns on in the status display lamp on the relevant channel.
7) The individual error contact and common error contact are activated.
When a system error occurs:
1) The power to the heaters of all detectors is removed.
2) The averaging outputs and all individual outputs are put into the status specified in
the Output Hold Setup display.
3) An error icon flashes in the switch indication area.
4) The red lamps turn on in the status display lamps on all channels.
5) The individual error contacts for all channels and common error contact are activated.