beautypg.com

Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 227

background image

IM 11M12D01-01E

12-17

12. Troubleshooting

12.3.2

Measured Value Lower Than True Value


(1) The measuring gas pressure becomes lower.

Where an increment of the measured value due to pressure change cannot be

neglected, take measures referring to subsection 12.3.1 (1).

(2) Moisture content in a reference gas changes (decreases) greatly.

If air at the detector installation site is used for the reference gas, large change of

moisture content in the air may cause an error in measured oxygen concentration
value (vol% O

2

).

When this error is not ignored, use a gas in which moisture content is constant such

as instrument air in almost dry condition as a reference gas.

In addition, change of moisture content in exhaust gas after combustion is also

considered as a cause of error. However, normally this error is negligible.

(3) Calibration gas (zero gas) is mixed into the detector due to leakage.

If the zero gas is mixed into the detector due to leakage as a result of failure of the

valve provided in the calibration gas tubing system, the measured value shows a
value a little lower than normal.

Check valves (needle valves, check valves, solenoid valves for automatic calibra-

tion, etc.) in the calibration gas tubing system for leakage. For manual valves, check
them after confirming that they are in fully closed states.

(4) Combustible components exist in the measured gas. If combustible components exist

in the measured gas, they burn in the sensor and thus oxygen concentration
decreases.
Check that there are no combustible components.

(5) Temperature of the detector cell reaches 750

Њ C or more.

12.3.3

Measurements Sometimes Show Abnormal Values


(1) Noise may be mixing in with the converter from the detector output wiring.

Check whether the converter and detector are securely grounded.
Check whether or not the signal wiring is laid along other power cords.

(2) The converter may be affected by noise from the power supply.

Check whether or not the converter power is supplied from the same outlet, switch,
or breaker as other power machines and equipment.

(3) Poor wiring contact

If there is poor contact in the wiring, the sensor voltage or thermocouple emf
(voltage) may vary due to vibration or other factors.

Check whether or not there are loose points in the wiring connections or loose
crimping (caulking) at the crimp-on terminal lugs.

(4) Combustible components in the measured gas may be getting into the sensor.

If the combustible components show signs of dust, the abnormality may be im-
proved by mounting a dust-removing filter K9471UA.

(5) There may be a crack in the sensor or leakage at the sensor-mounting portion.

If the indication of concentration varies in synchronization with the pressure change
in the furnace, check whether or not there is a crack in the sensor or whether the
sensor flange is sticking tightly to the probe-attaching face with the metal O-ring
squeezed.

(6) There may be leakage in the calibration gas tubing

In the case of a negative furnace inner pressure, if the indication of concentration
varies with the pressure change in the furnace, check whether or not there is leakage
in the calibration gas tubing.