2 calibration setup, Caution, Calibration setup – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 155: 1 mode
![background image](/manuals/360143/155/background.png)
IM 11M12D01-01E
9-5
9. Calibration
9.2 Calibration Setup
9.2.1 Mode
There are three calibration modes (for zero and span calibration): manual mode, semi-
auto mode (started from the display or started by contact input) using preset calibration
time and stabilization (settling) time, and autocalibration at preset intervals. Restrictions
related to each mode are described here.
- When [Manual Mode] is selected
In this mode, only Manual calibration can be performed. (Contact input used to start
semi-auto calibration has no effect). Autocalibration does not start at autocalibration
start time.
- When [Semi-Auto Mode] is selected.
In this mode, manual calibration or semi-auto calibration can be performed.
Autocalibration does not start at autocalibration start time.
- When [Auto Mode] is selected.
Any calibration mode is valid.
CAUTION
Indication Check mode of [Auto] cannot coexist with [Auto] Calibration mode. There-
fore, If the mode of Indication Check is set to [Auto], you cannot set the mode of
Calibration to [Auto], and vice versa.
< Setting Procedure>
1) From the basic panel display, if you touch the Setup key, the [Execution/Setup]
display appears. Select [Maintenance].
2) From the [Maintenance] display, select [Calibration setup] then [Mode] and you can
select Manual, Semi-Auto, or Auto in the selection pulldown that appears. ( Fig. 9.4
).
Enter
F52E.EPS
Mode:
᭜
Points:
᭜
Zero gas conc: 0 . 9 8 %
᭜
Span gas conc: 2 1 . 0 0 %
᭜
Timing
᭜
Cal. history initialize
Manual
Semi-Auto
Auto
Calibration setup
Fig 9.4 Calibration mode setting.