Calibration setup, 2 calibration setup procedure, 3 zero gas concentration – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 156: 4 span gas concentration

background image

IM 11M12D01-01E

9-6

9.2.2 Calibration Setup Procedure

You can perform both Span and Zero calibration, or either one or the other. Normally
select both: Span-Zero.


From [Maintenance] --> [Calibration setup] display, select [Points], then a selection
window opens and you can select Span-Zero, Span or Zero (Fig 9.5).

Enter

F9.5E.EPS

Calibration setup

Span
Zero

Span-Zero

Mode: Auto

Points:

Zero gas 9 8 %

Span gas conc: 2 1 . 0 0 %

Timing

Cal. history initialize

Fig. 9.5 Procedure for Setting Calibration Mode

9.2.3 Zero Gas Concentration

This sets the oxygen concentration of the zero gas used for calibration. Enter the value
from the label of the corresponding gas bottle.

9.2.4 Span Gas Concentration

This sets the oxygen concentration of the span gas used for calibration. When the
instrument air supply is used as the span gas, enter 21% O

2

.

Note

When the instrument air supply is to be used as span gas, cool it to -20

؇ C below dew

point to remove moisture, oil mist and dust from the air. If you do not do this to
purify the air, then the accuracy of the calibration may be affected.


1) From the [Calibration setup] display, select [Zero gas conc.], and the numerical data

entry display is displayed. Enter the oxygen concentration of the zero gas used for
calibration. For example, enter 0.98 vol% O

2

as [00098].

2) From the [Calibration setup] display, select [Span gas conc.], and the numerical data

entry display is displayed. Enter the oxygen concentration of the span gas used for
calibration. For example, enter 21 vol% O

2

as [02100].