beautypg.com

Semi-autocal, Semi-auto cal – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 161

background image

IM 11M12D01-01E

9-11

9. Calibration

Starting Semi-Auto Calibration from the Display
1) From the basic display display if you touch the Setup key, the [Execution/Setup]

display appears. Select [Calibration].

2) Next select [Semi-auto calibration]. The [Semi-autocal] channel select display,

Fig 9.8, is displayed.

3) Select the channel to be calibrated. The [Semi-auto cal.] display, Fig. 9.9, is dis-

played.

4) When you select [Start calibration] the [Semi-autocal] trend graph of Fig 9.10 is

displayed, and Span calibration begins.

5) When the preset calibration time has elapsed, Span calibration ends and Zero calibra-

tion begins. Likewise, when the preset calibration time has elapsed, Zero calibration
ends. After a preset stabilization (settling) time, the instrument reverts to measure-
ment mode.

6) Touch the Escape key (icon of door with arrow) to abort calibration. If you abort

Span or Zero calibration then calibration time is skipped and stabilization (settling)
time starts. If you touch the Escape key again then stabilization (settling) time is also
skipped, and you revert to the Basic display.

Enter

Semi-autocal.

Ch1

Ch2

Ch3

Ch4

Ch5

Ch6

Ch7

Ch8

F55E.EPS

Select calibration channel

Enter

F35E.EPS

Ch1

Semi-auto cal.

Check settings
Span gas conc: 2 1 . 0 0 %
Zero gas conc: 0 . 9 8 %
Cal. time: 2 min 0 0 s
Points: Span-Zero
Go to setup to change value.

Start calibration

Cancel calibration

Fig. 9.8 Channel Select Display Fig. 9.9 Semi-Auto Calibration Start.

F9.10E.EPS

21.00%

Ch1

1min./div.

0.98%

21.0 %O2

CAL. TIME

Semi-autocal.

Fig. 9.10 Semi-Auto Calibration Trend Graph