Semi-autocal, Semi-auto cal – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 161
IM 11M12D01-01E
9-11
9. Calibration
Starting Semi-Auto Calibration from the Display
1) From the basic display display if you touch the Setup key, the [Execution/Setup]
display appears. Select [Calibration].
2) Next select [Semi-auto calibration]. The [Semi-autocal] channel select display,
Fig 9.8, is displayed.
3) Select the channel to be calibrated. The [Semi-auto cal.] display, Fig. 9.9, is dis-
played.
4) When you select [Start calibration] the [Semi-autocal] trend graph of Fig 9.10 is
displayed, and Span calibration begins.
5) When the preset calibration time has elapsed, Span calibration ends and Zero calibra-
tion begins. Likewise, when the preset calibration time has elapsed, Zero calibration
ends. After a preset stabilization (settling) time, the instrument reverts to measure-
ment mode.
6) Touch the Escape key (icon of door with arrow) to abort calibration. If you abort
Span or Zero calibration then calibration time is skipped and stabilization (settling)
time starts. If you touch the Escape key again then stabilization (settling) time is also
skipped, and you revert to the Basic display.
Enter
Semi-autocal.
Ch1
᭜
Ch2
᭜
Ch3
᭜
Ch4
᭜
Ch5
᭜
Ch6
᭜
Ch7
᭜
Ch8
F55E.EPS
Select calibration channel
Enter
F35E.EPS
Ch1
Semi-auto cal.
Check settings
Span gas conc: 2 1 . 0 0 %
Zero gas conc: 0 . 9 8 %
Cal. time: 2 min 0 0 s
Points: Span-Zero
Go to setup to change value.
Start calibration
᭜
Cancel calibration
Fig. 9.8 Channel Select Display Fig. 9.9 Semi-Auto Calibration Start.
F9.10E.EPS
21.00%
Ch1
1min./div.
0.98%
21.0 %O2
CAL. TIME
Semi-autocal.
Fig. 9.10 Semi-Auto Calibration Trend Graph