7 parameter initialization, Warning, Defaults – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 189
IM 11M12D01-01E
10-27
10. Other Functions
10.7 Parameter Initialization
Parameter settings can be initialized to the factory default settings. Initialization can be
done for all parameters or for individual parameters. The parameters that can be initial-
ized and their defaults are listed in Table 10.7.
display.
2) Select [Commissioning], next [Others] then [Defaults]. A display like Figure 10.32.1
appears.
3) Select the desired item to be initialized then a display like Figure 10.32.3 appears.
Select [Defaults start] then initialization starts.
4) When initialization completes then you are returned to a display like Figure 10.32.2.
Display
᭜
Calibration
᭜
Blow back data
᭜
mA-outputs
᭜
Alarms
᭜
Contacts data
᭜
Others
᭜
All data
Enter
Defaults
F10.32.1E.EPS
Fig. 10.32.1 Defaults Display
Enter
F10.32.2E.EPS
᭜
Defaults start
Defaults
All data
Cancel
Fig. 10.32.2 "Initialize All Data" Start Display
WARNING
Do not turn off power during initialization. Otherwise, initialization will not be per-
formed properly