Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

80

6.7.2.5. Editing DAS Parameters

Data channels can be edited individually from the front panel without affecting other data channels.
However, when editing a data channel, such as during adding, deleting or editing parameters, all data for
that particular channel will be lost, because the DAS can store only data of one format (number of
parameter columns etc.) for any given channel. In addition, an DAS configuration can only be uploaded
remotely as an entire set of channels. Hence, remote update of the DAS will always delete all current
channels and stored data.

To modify, add or delete a parameter, follow the instruction shown in Section 6.7.2.2 then press:


Edit Data Parameter Menu

Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 1, 900


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC

SET>

EDIT

PRINT

EXIT

SETUP X.X

PARAMETERS:1

EDIT

PRINT

EXIT

SETUP X.X

0) PARAM=CONC1, MODE=AVG

PREV NEXT

INS DEL EDIT EXIT

Press SET> button until…

SETUP X.X

EDIT PARAMS (DELETE DATA)

YES NO

NO returns to

the previous

menu and

retains all data.

Moves the

display between

existing

Parameters

Inserts a new Parameter

before the currently

displayed Parameter

Deletes the Parameter

currently displayed.

Use to configure

the functions for

this Parameter.

Exits to the main

Data Acquisition

menu

YES will delete

all data in that

entire channel.

From the DATA ACQUISITION menu