beautypg.com

Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 245

background image

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

227

Table 10-5: Relay Board Status LED Failure Indications

LED FUNCTION

SIGNAL I/O PARAMETER

DIAGNOSTIC TECHNIQUE

ACTIVATED BY

VIEW RESULT

D2

Yellow

WHEEL

HEATER

WHEEL_HEATER WHEEL_TEMP

Voltage displayed should change. If not:
Failed Heater
Faulty Temperature Sensor
Failed AC Relay
Faulty Connectors/Wiring

D3

Yellow

BENCH

HEATER

BENCH_HEATER BENCH_TEMP

Voltage displayed should change. If not:
Failed Heater
Faulty Temperature Sensor
Failed AC Relay
Faulty Connectors/Wiring

D4

Yellow

SPARE

N/A N/A

N/A

D5

Green

SAMPLE/CAL

GAS VALVE

OPTION

CAL_VALVE N/A

Sample/Cal Valve should audibly change states. If
not:
Failed Valve
Failed Relay Drive IC on Relay Board
Failed Relay Board
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring

D6

Green

ZERO/SPAN

GAS VALVE

OPTION

SPAN_VALVE N/A

Zero/Span Valve should audibly change states. If
not:
Failed Valve
Failed Relay Drive IC on Relay Board
Failed Relay Board
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring

D7

Green

SHUTOFF

VALVE OPTION

SHUTOFF_VALVE N/A

Shutoff Valve should audibly change states. If not:
Failed Valve
Failed Relay Drive IC on Relay Board
Failed Relay Board
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring

D8

Green

IR SOURCE

IR_SOURCE CO

2

_MEASURE

Voltage displayed should change. If not:
Failed IR Source
Faulty +12 VDC Supply (PS2)
Failed Relay Board
Failed IR Photo-Detector
Failed Sync/Demod Board
Faulty Connectors/Wiring