beautypg.com

Connecting the status outputs – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 41

background image

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

23

3.5.4. Connecting the Status Outputs

If you wish utilize the analyzer’s status outputs to interface with a device that accepts logic-level digital
inputs, such as programmable logic controllers (PLC’s) they are accessed via a 12-pin connector on the
analyzer’s rear panel labeled STATUS.

STATUS

1 2 3 4 5 6 7 8 D

+

NOTE

Most PLC’s have internal provisions for limiting the current the input will draw.

When connecting to a unit that does not have this feature, external resistors must

be used to limit the current through the individual transistor outputs to ≤50mA

(120 Ω for 5V supply).

The pin assignments for the status outputs can be found in the table below:

Table 3-5: Status Output Pin-outs

OUTPUT #

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

On if no faults are present.

2

CONC VALID

On if CO

2

concentration measurement is valid.

If the CO

2

concentration measurement is invalid, this bit is OFF.

3

HIGH RANGE

On if unit is in high range of DUAL or AUTO range modes.

4

ZERO CAL

On whenever the instruments ZERO point is being calibrated.

5

SPAN CAL

On whenever the instruments SPAN point is being calibrated.

6

DIAG MODE

On whenever the instrument is in DIAGNOSTIC mode.

7

ALARM1

On whenever the measured CO

2

concentration is above the set point for

ALM1

8

ALARM2

On whenever the measured CO

2

concentration is above the set point for

ALM2

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bused together.

+

DC POWER

+ 5 VDC

Digital Ground

The ground level from the analyzer’s internal DC power supplies.