Analog output step test – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 140
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
122
To enter the signal I/O test mode, press:
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
EXAMPLE
SAMPLE
ENTER SETUP PASS: 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
PREV NEXT
JUMP
ENTR
EXIT
DIAG I / O
Test Signals Displayed Here
PREV NEXT
JUMP
PRNT EXIT
Use JUMP to go
directly to a specific
signal location
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O
JUMP TO: 12
1 2
ENTR
EXIT
EXAMPLE:
Enter 12 to Jump to
12)
ST_CONC_VALID
DIAG I / O
ST_CONC_VALID = ON
PREV NEXT
JUMP
ON
PRNT
EXIT
Exit to return
to the
DIAG
menu
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM
EXIT
Use NEXT & PREV to
move between signal
types.
Pressing PRNT will send a formatted printout to the serial port and can be captured
with a computer or other output device.
EXIT
returns
to the main
SAMPLE
display
6.13.3. Analog Output Step Test
This test can be used to check the accuracy and proper operation of the analog outputs. The test forces
all four analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20%
increments. This test is useful to verify the operation of the data logging/recording devices attached to
the analyzer.