Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

133


DIAG AIO

CONC_OUT_2 SPAN: 10000 mV

U100 UP10

UP DOWN DN10

D100

ENTR EXIT

DIAG AIO

AUTO CALIBRATING CONC_OUT_2

DIAG AIO

AIN CALIBRATED:

NO

SET>

EDIT

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

FROM ANALOG I/O CONFIGURATION MENU

DIAG AIO

CONC_OUT_2 RANGE: CURR

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 CALIBRATED: NO

< SET

CAL

EXIT

DIAG AIO

CONC_OUT_2:CURR, NO CAL

< SET SET>

EDIT

EXIT

Press SET> to select the analog output channel

to be configured:.

Increase or decrease the current

output by 100, 10 or 1 counts.

The resulting change in output

voltage is displayed in the upper

line.

Continue adjustments until the

correct current is measured with

the current meter.

DIAG AIO

CONC_OUT_2 ZERO: 27 mV

U100 UP10

UP DOWN DN10 D100

ENTR EXIT

EXIT

ignores the

new setting, ENTR

accepts the new

setting.

DIAG AIO

CONC_OUT_2 CALIBRATED: YES

< SET

CAL

EXIT

DIAG AIO

CONC_OUT_2 ZERO: 9731 mV

U100 UP10

UP DOWN DN10 D100

ENTR EXIT

DIAG AIO

CONC_OUT_2 CURR-Z: 0 mV

U100

UP10

UP DOWN DN10 D100

ENTR EXIT

The instrument attempt to automatically calibrate

the channel … then beep.

DIAG AIO

CONC_OUT_2 D/A/ CAL ERROR

EXIT

If a current meter is not available, an alternative method for calibrating the current loop outputs is to
connect a 250

  1% resistor across the current loop output. Using a voltmeter connected across the

resistor, follow the procedure above but adjust the output to the following values:

Table 6-21: Current Loop Output Calibration with Resistor

Full scale

Voltage for 2-20 mA

(measured across resistor)

Voltage for 4-20 mA

(measured across resistor)

0%

0.5 V

1.0 V

100%

5.0 V

5.0 V