Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
129
To select manual output calibration for a particular channel, activate the ANALOG I/O
CONFIGURATION MENU (See Section 6.13.1), then press:
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
ON ENTR EXIT
Exit to return to
the main
sample display
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
SET>
CAL
EXIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
Toggles the
auto cal mode
ON/ OFF for
this analog
output channel
only.
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
Press SET> to select the analog output channel to
be configured. Then press EDIT to continue
ENTR
accepts the new setting
and returns to the previous
menu. EXIT ignores the new
setting and returns to the
previous menu.
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
< SET
SET> EDIT
EXIT
Now the analog output channels should either be automatically calibrated or they should be set to manual
calibration, which is described next.
6.13.4.3. Manual Analog Output Calibration and Voltage Adjustment
For highest accuracy, the voltages of the analog outputs can be manually calibrated. Calibration is done
through the instrument software with a voltmeter connected across the output terminals (Figure 6-15).
Adjustments are made using the touchscreen buttons by setting the zero-point first and then the span-
point (Table 6-20).
The software allows this adjustment to be made in 100, 10 or 1 count increments.
Table 6-20: Voltage Tolerances for Analog Output Calibration
Full Scale
Zero Tolerance
Span Voltage
Span Tolerance
0.1 VDC
±0.0005V
90 mV
±0.001V
1 VDC
±0.001V
900 mV
±0.001V
5 VDC
±0.002V
4500 mV
±0.003V
10 VDC
±0.004V
4500 mV
±0.006V
NOTE
Outputs configured for 0.1V full scale should always be calibrated manually