beautypg.com

Purge gas pressure control, Particulate filter, Pneumatic sensors – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 214

background image

Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

196

9.2.3. Purge Gas Pressure Control

In order to ensure that all of the ambient CO

2

is purged from the GFC Wheel housing an adequate supply

of dried air, scrubbed of CO

2

must be supplied to the PURGE AIR inlet at the back of the instrument.

 The minimum gas pressure of the source of purge air should be 7.5 psig.

 If the source of the purge air is shared by a Teledyne TAI’s T700 (as shown in figure 3-7) the

minimum gas pressure should be 25 psig and should not exceed 35 psig.

In order to maintain the proper pressure differential between the inside of the GFC wheel housing and
ambient air, the GFC 7000TA design includes a manually adjustable pressure regulator (see Figure 3-5
and Figure 3-6 for

Purge Gas Pressure Regulator

) that maintains the pressure of the purge air feed at 7.5

psig.

9.2.4. Particulate Filter

The Model GFC 7000TA Analyzer comes equipped with a 47 mm diameter, Teflon, particulate filter with a
5 micron pore size. The filter is accessible through the front panel, which folds down to allow access, and
should be changed according to the suggested maintenance schedule described in Table 9-1.

9.2.5. Pneumatic Sensors

9.2.5.1. Sample Pressure Sensor

An absolute value pressure transducer plumbed to the outlet of the sample chamber is used to measure
sample pressure. The output of the sensor is used to compensate the concentration measurement for
changes in air pressure. This sensor is mounted to a printed circuit board with the sample flow sensor on
the sample chamber; see the following section and Figure 3-3.

9.2.5.2. Sample Flow Sensor

A thermal-mass flow sensor is used to measure the sample flow through the analyzer. The sensor is
calibrated at the factory with ambient air or N

2

, but can be calibrated to operate with samples consisting of

other gases such as CO

2

, See Section 9.3.4. This sensor is mounted to a printed circuit board with the

Sample Pressure sensor on the sample chamber; see the previous section and Figure 3-3.