Pressure calibration – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 156

Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
138
6.13.7. Pressure Calibration
A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas. This
data is used to compensate the final CO
2
concentration calculation for changes in atmospheric pressure
and is stored in the CPU’s memory as the test function PRES (also viewable via the front panel).
NOTE
This calibration must be performed when the pressure of the sample gas is equal
to ambient atmospheric pressure.
Before performing the following pressure calibration procedure, disconnect the
sample gas pump and the sample gas-line vent from the sample gas inlet on the
instrument’s rear panel.
To cause the analyzer to measure and record a value for PRES, press.
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM
EXIT
Adjust these values until the
displayed pressure equals the
pressure measured by the
independent pressure meter.
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit at
any time
to return
to main
the
SETUP
menu
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
Repeat Pressing NEXT until . .
.
DIAG PCAL
ACTUAL PRESS : 27.20 IN-HG-A
2 7 .2 0 ENTR EXIT
ENTR accepts the
new value and
returns to the
previous menu
EXIT
ignores the
new value and
returns to the
previous menu