Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

65

6.2.1. Test Functions

Test functions are shown on the front panel display while the analyzer is in SAMPLE mode. They provide
information about the present operating status of the instrument and can be used troubleshooting and can
be recorded in one of the DAS channels (Section 6.7) for data analysis. To view, press the buttons.

Table 6-2: Test Functions Defined

PARAMETER DISPLAY

UNITS

MEANING

Range

RANGE

RANGE1

1

RANGE2

1

PPB, PPM,

UGM, MGM

The full scale limit at which the reporting range of the analyzer is
currently set. THIS IS NOT the Physical Range of the instrument.
See Section 6.8.1 for more information.

O

2

Range

O2 RANGE

%

The range setting for the O

2

Sensor

Stability

STABIL

PPB, PPM

UGM, MGM

Standard deviation of CO

2

concentration readings. Data points are

recorded every ten seconds using the last 25 data points.

CO

2

Measure

MEAS

MV

The demodulated, peak IR detector output during the measure
portion of the GFC Wheel cycle.

CO

2

Reference

REF

MV

The demodulated, peak IR detector output during the reference
portion of the GFC wheel cycle.

Measurement /

Reference Ratio

MR Ratio

-

The result of CO2 MEAS divided by CO2 REF. This ratio is the
primary value used to compute CO

2

concentration. The value

displayed is not linearized.

Sample Pressure

PRES

In-Hg-A

The absolute pressure of the Sample gas as measured by a
pressure sensor located inside the sample chamber.

Sample Flow

SAMPLE FL

cc/min

Sample mass flow rate. This is computed from the differential
between the pressures measured up-stream and down-stream of the
sample critical flow orifice pressures.

Sample

Temperature

SAMP TEMP

C

The temperature of the gas inside the sample chamber.

Bench

Temperature

BENCH TEMP

C

Optical bench temperature.

Wheel

Temperature

WHEEL TEMP

C

Filter wheel temperature.

Box Temperature

BOX TEMP

C

The temperature inside the analyzer chassis.

O

2

Cell

Temperature

2

O2 CELL

TEMP

2

C

O

2

sensor cell temperature.

Photo-detector
Temp. Control

Voltage

PHT DRIVE

mV

The drive voltage being supplied to the thermoelectric coolers of the
IR photo-detector by the sync/demod Board.

Slope

SLOPE

-

The sensitivity of the instrument as calculated during the last
calibration activity. The SLOPE parameter is used to set the span
calibration point of the analyzer.

Offset

OFFSET

-

The overall offset of the instrument as calculated during the last
calibration activity. The OFFSET parameter is used to set the zero
point of the analyzer response.

O2 Sensor Slope

2

O2 SLOPE

- O

2

slope, computed during zero/span calibration.

O2 Sensor Offset

2

O2 OFFSET

- O

2

offset, computed during zero/span calibration.

Test channel
output signal

TEST

mV, mA

Displays the signal level of the TEST analog output channel. Only
appears when the TEST channel has been activated.

Current Time

TIME

-

The current time. This is used to create a time stamp on DAS
readings, and by the AUTOCAL feature to trigger calibration events.

1

Only appears when the instrument’s reporting range mode is set for DUAL or AUTO

2

Only appears when the optional O

2

sensor is installed.