beautypg.com

Auto range mode (auto) – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 112

background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

94

6.8.5. Auto Range Mode (AUTO)

In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined
ranges (low and high). The unit will switch from low range to high range when the CO

2

concentration

exceeds 98% of the low range span. The unit will return from high range back to low range once both the
CO

2

concentration falls below 75% of the low range span.

In AUTO Range mode the instrument reports the same data in the same range on both the A1 and A2
outputs and automatically switches both outputs between ranges as described above. Also, the RANGE
test function displayed on the front panel will automatically switch to show which range is in effect.

The high/low range status is also reported through the external, digital status outputs (Section 6.15.1.1).

To set individual ranges press the following keystroke sequence.


SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE:

SNGL

SNGL DUAL

AUTO

ENTR EXIT

EXIT

x 2 returns

to the main

SAMPLE

display

SETUP X.X

RANGE MODE:

AUTO

SNGL

DUAL AUTO

ENTR

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

HIGH RANGE:

500.0 Conc

0 0 5 0 0 .0

ENTR

EXIT

SETUP X.X

LOW RANGE:

500.0 Conc

0 0 5 0 0 .0

ENTR

EXIT

Toggle the numeral

buttons to set the

LOW

and HIGH

range value.

ENTR accepts the

new setting, EXIT

ignores the new

setting.

CAUTION

In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets

for computing CO2 concentration.

The two ranges must be independently calibrated.

NOTE

Avoid accidentally setting the low range of the instrument with a higher span limit

than the high range. This will cause the unit to stay in the low reporting range

perpetually and defeat the function of the AUTO range mode.