6.12. SETUP

MORE VARS: Internal Variables (VARS)

The GFC 7000TA has several-user adjustable software variables, which define certain operational
parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be
manually re-defined using the VARS menu. Table 6-14 lists all variables that are available within the 818
password protected level.

Table 6-14: Variable Names (VARS) Revision B.3

NO. VARIABLE

DESCRIPTION

ALLOWED

VALUES

0 DAS_HOLD_OFF

Changes the internal data acquisition system (DAS) hold-off
time, which is the duration when data are not stored in the
DAS because the software considers the data to be
questionable. That is the case during warm-up or just after
the instrument returns from one of its calibration modes to
SAMPLE mode. DAS_HOLD_OFF can be disabled entirely
in each DAS channel.

Can be between 0.5

and 20 minutes

Default=15 min.

1 CONC_PRECISION

Allows the user to set the number of significant digits to the
right of the decimal point display of concentration and
stability values.

AUTO, 1, 2, 3, 4

Default=AUTO

2

STABIL_GAS

Selects which gas measurement is displayed when the STABIL
test function is selected.

CO

2

& O

2

1

3 DYN_ZERO

Dynamic zero automatically adjusts offset and slope of the
CO

2

response when performing a zero point calibration

during an AutoCal (Section 7).

ON/OFF

4 DYN_SPAN

Dynamic span automatically adjusts slope and slope of the
CO

2

response when performing a zero point calibration

during an AutoCal (Section 7).

Note that the DYN_ZERO and DYN_SPAN features are not
allowed for applications requiring EPA equivalency.

ON/OFF

5 CLOCK_ADJ

Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock is
too fast.

-60 to +60 s/day

1

O

2

option is only available in analyzers with O

2

sensor options installed.