beautypg.com

Step 2 – activate o, Concentration to the o, Sensor output, press – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 59

background image

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

41

SAMPLE

GAS TO CAL:CO2

CO2

O2

ENTR EXIT

SAMPLE RANGE = 500.000 PPM CO2 =XXX.X

< TST TST > CAL

SETUP

M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX

ZERO SPAN

CONC

EXIT

SAMPLE

GAS TO CAL:O2

NOX

O2

ENTR

EXIT

M-P CAL O2 SPAN CONC:20.8%

0

2

0

.8

0

ENTR EXIT

The O

2

span concentration value automatically defaults to

20.8 %

.

If this is not the the concentration of the span gas being

used, toggle these buttons to set the correct concentration

of the O

2

calibration gases.

EXIT

ignores the new

setting and returns to

the previous display.

ENTR

accepts the new

setting and returns to

the previous menu.

STEP 2 – ACTIVATE O

2

SENSOR STABILITY FUNCTION

To change the stability test function from NO

x

concentration to the O

2

sensor output, press: